Property | Description/Value |
---|---|
Dew point | |
Sensor | DRYCAP® 180S |
Measurement range | −40 … +100 °C (−40 … +212 °F) Td/f |
Accuracy |
±2 °C (±3.6 °F) Td/f See accuracy graph |
Response time 63 % [90 %] 1 | |
From dry to wet | 5 s [10 s] |
From wet to dry | 45 s [5 min] |
Temperature | |
Measurement range | 0 … +180 °C (+32 … +356 °F) |
Accuracy at +100 °C (+212 °F) | ±0.4 °C (±0.72 °F) |
Temperature sensor | Pt100 RTD Class F0.1 IEC 60751 |
Mixing ratio | |
Measurement range (typical) | 0–1000 g/kg (0–7000 gr/lbs) |
Accuracy (typical) | ±12 % of reading |
Absolute humidity | |
Measurement range | 0–600 g/m3 |
Accuracy | ±10 % of reading (typical) |
Property | Description/Value |
---|---|
Operating temperature range for probe head | −40 … +180 °C (−40 … +356 °F) |
Operating temperature range for probe body | −40 … +80 °C (−40 … +176 °F) |
Storage temperature |
−40 … +80 °C (−40 … +176 °F) |
Measurement environment |
For air, nitrogen, hydrogen, argon, helium, and oxygen 2 |
IP rating for probe body |
IP66: Dust-tight. Protected from powerful water jets from any direction. |
Property | Description/Value |
---|---|
Operating voltage | 15–30 V DC |
Current consumption | 10 mA typical, 500 mA max. |
Digital output | RS-485, non-isolated |
Protocols | Modbus RTU |
Property | Description/Value |
---|---|
Connector | M12 5-pin A-coded male |
Weight (with a 2-m cable) | 436 g (15.37 oz) |
Probe cable length | 2 m (6.56 ft) or 10 m (32.8 ft) |
Materials | |
Probe | AISI 316L |
Probe body | AISI 316L |
Cable jacket | FEP |
1 Tested with
sintered filter.
2
Consult Vaisala if other chemicals are present. Consider safety regulations with flammable gases.