Refractive index nD verification - PR-21-S

PR-21 Series User Guide

Document code
IM-EN-PR21S
Revision
C
Language
English
Product
PR-21-S
Document type
User guide

Before starting the verifications procedures, make sure that you have a PR‑21‑S sample holder at hand. Check the condition of your standard refractive index liquids. You need a cleaning solution (ethanol) to clean the sensor prism and the sample holder.

The sample holder is kept in place by a rubber link.

Figure 1. Sample holder: Part number PR‑5002

The verification of the PR‑21‑S sensor calibration is made using a set of standard refractive index liquids with the nominal values at +25 °C (+77 °F):

  • 1.330
  • 1.350
  • 1.360
  • 1.370
  • 1.390
  • 1.410
  • 1.430
  • 1.450
  • 1.460
  • 1.470
  • 1.490
  • 1.500
  • 1.520

The accuracy of the certified standard refractive index liquids is ±0.0002 and they can be traced back to national standards: N.I.S.T Standards # 1823 and # 1823 II.

The repeatability of PR‑21‑S sensor, that is the deviation from the latest nD calibration, is within ±0.0004.

As the specified accuracy of PR‑21‑S is ±0.0004, then the representative level is the root sum square of the three accuracy specifications, that is ±0.0006.

Vaisala provides a set of standard R.I. liquids, PR‑2000, containing these five liquids. The set can be ordered directly from Vaisala or through your nearest representative.