Linear damping - PR-21-S

PR-21 Series User Guide

Document code
IM-EN-PR21S
Revision
C
Language
English
Product
PR-21-S
Document type
User guide

If the process has fast step changes, linear (fast) damping gives shorter settling time. In the linear damping (fast damping), the output is the running average of the signal during the damping time. After a step change the signal rises linearly and reaches the final value after the damping time. The linear damping gives the best trade-off between random noise suppression and step change response time. Use the 4 DAMPING TIME menu item to set the damping time.

For similar noise suppression a longer damping time has to be specified than for the exponential damping.

The following figure shows how linear damping time affects the measurement.

Figure 1. Linear damping