Equipment - PR-21-S

PR-21 Series User Guide

Document code
IM-EN-PR21S
Revision
C
Language
English
Product
PR-21-S
Document type
User guide

The intrinsically safe process refractometer consists of the following:

  • A modified refractometer sensor PR‑21‑S‑…‑IA
  • An indicating transmitter STR with single sensor connectivity
  • IS isolator and cabling between the refractometer sensor and the transmitter
Figure 1. Refractometer system PR‑21‑S‑...‑IA

The equipment is intrinsically safe only if all mounting instructions are followed, see Intrisincally safe mounting. If the instrument has been in any way damaged during transportation, return it to your nearest Vaisala K‑PATENTS® service point for checkup before installation. Never install a damaged instrument into the process line.

The intrinsically safe sensor PR‑21‑S‑…‑IA is identified by the nameplate, see the following. The indicating transmitter is of model STR, for single sensor connection.

Figure 2. IA nameplate

An intrinsically safe sensor has a different processor card and a different terminator card than a standard sensor, other parts are as in standard sensor (see earlier in this chapter for full parts list).

Figure 3. Intrinsically safe parts
WARNING Do not replace any part of an intrinsically safe sensor with a standard sensor part.
Figure 4. Warning sticker
WARNING If the sensor cover is made of aluminum, the refractometer sensor can cause ignition if it hits other metal parts during the installation. An aluminum sensor cover must have a sticker warning about this possibility.