Diagnostic message priorities - PR-21-S

PR-21 Series User Guide

Document code
IM-EN-PR21S
Revision
C
Language
English
Product
PR-21-S
Document type
User guide
The messages are listed in descending order of priority. For example, if both NO OPTICAL IMAGE and TEMP MEASUREMENT FAULT are activated, only NO OPTICAL IMAGE displays.

When a relay is configured with FUNCTION INSTRUMENT OK, see Configuring relays, it is closed when there is no equipment malfunction.

Certain malfunctions cause the mA measurement to return to the mA default output value, see Configuring mA outputs. For more information, see the following table.

Message Instrument OK Returns to default mA
Conc Temp
SHORT-CIRCUIT X X
NO SIGNAL X X
OUTSIDE LIGHT ERROR
NO OPTICAL IMAGE X
TEMP MEASUREMENT FAULT X X
PRECONDITIONING X
WASH X
RECOVERING X
HIGH SENSOR HUMIDITY
HIGH SENSOR TEMP
HIGH TRANSMITTER TEMP
LOW TRANSMITTER VOLT
EXTERNAL WASH STOP X
LOW TEMP WASH STOP X
NO SAMPLE / WASH STOP X
EXTERNAL HOLD X
NO SAMPLE X X
PRISM COATED X X
OUTSIDE LIGHT TO PRISM X
LOW IMAGE QUALITY X
PRISM WASH FAILURE X
PRISM WASH WARNING X
NO SENSOR X X
NORMAL OPERATION X