Indicating transmitter specifications - PR-21-S

PR-21 Series User Guide

Document code
IM-EN-PR21S
Revision
C
Language
English
Product
PR-21-S
Document type
User guide
Table 1. Indicating transmitter specifications
Feature Specification
Display 320 x 240 pixel graphical LCD with LED backlight
Keypad

18 membrane keys

Current output Two independent current sources, 4 … 20 mA, max. load 1000 Ω, galvanic isolation 1500 V DC or AC (peak), hold function during prism wash.
AC power supply AC input 100 … 240 V AC, ±10%, 50/60 Hz / 30 VA, fuse voltage 250 V AC, fuse max. size 10 A, fuse speed slow.
DC power supply DC input 24 V DC, ±10%, fuse voltage min. 48 V DC, fuse max. size 4 A, fuse speed fast.
Overvoltage category Overvoltage category II
Alarms/Wash relays Two built-in signal relays, max. 240 V/3 A, fuse with voltage 250 V AC, max. size 10 A and speed slows.
Input switches Four switch inputs
Current outputs

Two current outputs configurable independently to indicate process concentration or temperature of either sensor.

Sensor connectivity, DTR

One or two sensors can be connected to the transmitter. Sensors independent of each other: own parameter sets and usable in different applications.

Sensor connectivity, STR

Only one sensor can be connected to the transmitter. Used with the Intrinsically safe sensor PR‑23‑…‑IA/‑IF/‑CI.

Transmitter protection class

Enclosure IP66, Nema 4X

Pollution degree Pollution degree 2
Installation

Indoor

Indicating transmitter weight

4.5 kg (10 lbs)

Altitude Max. altitude 2000 m (6561.68 ft)
Ambient temperature 0 … +45 °C (+32 … +113 °F)

Display operating temperature 0 … +50 °C (+32 … +122 °F), display storage temperature ‑20 … +60 °C (‑4 … +140 °F).

Relative humidity 0 … 100 %
Wet location Not applicable