The quality control (QC) parameters
determine the limits for certain observations. The quality control limits are optional. If a
limit is missing for a measurement, the corresponding check is omitted.
Do not change the values unless you know what you are doing.
In the Web UI, go to Settings > Station parameters.
In the qc parameter group, edit the values that you want.
The parameters are as follows:
lo: Lower limit
for the measurement
hi: Upper limit
for the measurement
over: The
allowed overshoot as percentage of the full measurement range (hi
– lo). If the measurement value exceeds the lower or upper limit
but stays within the overshoot value, the measurement is still considered valid but
its value is set to the respective limit value.
step: Maximum
step between 2 successive measurements
To save the changes and take them into use, select Save and restart
script.