Complete the steps listed in Adjusting RH and T with MI70 handheld indicator before continuing with this adjustment option. |
This adjustment is made using relative humidity references 11.3 % RH (LiCl) and 75.5 % RH (NaCl). |
- Remove the filter from the transmitter's probe and insert the probe head into the LiCl salt chamber.
- Press ADJUST to continue adjusting.
- Choose LiCl-NaCl autom. from the MI70 adjustment menu and press SELECT. Press OK to accept the note telling about references.
- Press READY when the reading has stabilized in the LiCl salt chamber (can take 30 minutes or more). You can follow the stabilization from the GRAPH display.
- Remove the probe from the LiCl salt chamber and insert the probe head into the NaCl salt chamber.
- Press READY when the reading has stabilized in the NaCl salt chamber (can take 30 minutes or more). You can follow the stabilization from the GRAPH display.
- Confirm the adjustment by pressing YES (by pressing NO you return to adjustment mode display and no changes are made).
- The adjustment is done. Press BACK and EXIT to return to the basic display.
- Turn off the MI70 and detach the connection cable.