Probe Mounting Using the Duct Installation Kit - 210697 - HMP3 - HMP7 - HMT330 - SWINSIGHT10

Duct Installation Kit 210697 Installation Guide

Document code
M212351EN
Revision
A
Language
English
Product
210697
HMP3
HMP7
HMT330
Document type
User guide
  1. Thread the clamping bush, nut, and supporting bar onto the probe cable.
    1
    M9×1 thread on probe head
    2
    Supporting bar
    3
    Clamping bush
    4
    Nut
  2. Thread the hanger to the probe and then thread the supporting bar to the hanger. Fasten the bushing around the cable and the supporting bar.
    1
    Hanger
    2
    Bushing
  3. Pass the probe through the flange, and thread the nut to the flange.
    1
    Flange
    2
    Clamping bush
    3
    Bushing
    4
    Nut
  4. Fasten the seal to the flange.
    1
    Seal
    2
    Flange
    3
    Plastic washer, 4 pcs
    4
    Washer, 4 pcs
    5
    Screw, 4 pcs
  5. Make a Ø 20 mm hole in the surface of the chamber or duct.
  6. Insert the probe through the hole and fasten the flange using screws and washers.