Probe Mounting Using the Duct Installation Kit - 210697 - HMP3 - HMP7 - HMT330 - SWINSIGHT10
Duct Installation Kit 210697 Installation Guide
- Document code
- M212351EN
- Revision
- A
- Language
- English
- Product
- 210697
- HMP3
- HMP7
- HMT330
- Document type
- User guide
-
Thread the clamping bush, nut, and supporting bar onto the probe cable.
- 1
- M9×1 thread on probe head
- 2
- Supporting bar
- 3
- Clamping bush
- 4
- Nut
-
Thread the hanger to the probe and then thread the supporting bar to the hanger. Fasten
the bushing around the cable and the supporting bar.
- 1
- Hanger
- 2
- Bushing
-
Pass the probe through the flange, and thread the nut to the flange.
- 1
- Flange
- 2
- Clamping bush
- 3
- Bushing
- 4
- Nut
-
Fasten the seal to the flange.
- 1
- Seal
- 2
- Flange
- 3
- Plastic washer, 4 pcs
- 4
- Washer, 4 pcs
- 5
- Screw, 4 pcs
-
Make a Ø 20 mm hole in the surface of the chamber or duct.
-
Insert the probe through the hole and fasten the flange using screws and washers.