2-point T adjustment of HMP70 series probes - HM70

HM70 User Guide

Document code
M210297EN
Revision
G
Language
English
Product
HM70
Document type
User guide
Carry out the steps listed in Making temperature adjustments before continuing with this adjustment option.
The difference between the 2 references must be at least 30 °C or the adjustment cannot be done.
To complete the T adjustment using the 2-point adjustment method:
  1. Choose 2-point adjustment and press Select.
  2. Set the probe to the lower reference temperature. You can follow the stabilization from the graphical display by pressing Graph. Press Ready when the reading has stabilized.
  3. Enter the correct lower reference temperature value with the arrow buttons and press OK.
  4. Set the probe to the higher reference temperature. You can follow the stabilization from the graphical display by pressing Graph. Press Ready when the reading has stabilized.
  5. Enter the correct higher reference temperature value with the arrow buttons and press OK.
  6. Confirm the adjustment by pressing Yes. If you press No, you return to the adjustment mode display and no changes are made.
  7. The adjustment is done. Press Back and Exit to return to the basic display.