You can check and adjust the humidity and temperature channels of HMT330 and PTU300 with the HM70 meter. If you adjust both channels, make the
temperature adjustment first. Calibrate your transmitter against a calibrated HMP70 series reference probe or against a calibrator's reference humidity by using the MI70 indicator in
communication.
To connect HM70 to HMT330/PTU300 transmitters, you need the optional calibration cable (Vaisala item code 211339).
First carry out steps 1 ... 9. After 9, choose 1 of the alternative
methods for completing the adjustment.
To start the adjustment:
Connect calibration cable 211339 to the SERVICE PORT connector on the HMT330/PTU300 component board.
Connect the other end of the cable to either one of the connector ports located on the bottom of the MI70 indicator. See Figure 1.
Switch on both devices.
The reading of the transmitter is shown on the first or middle row of the display, depending on the connector port to which the calibration cable is connected. The value of port I is
shown on the upper row of the display and the value of port II on the lower row of the display. The difference of the readings is shown on the lowest row.
If you are using an HMT330 unit with the optional chemical purge function, activate the chemical purge before calibration and adjustment. To activate the purge with the MI70
indicator:
This step applies only to HMT330 series transmitters.
Open Menu by pressing
Open
Select Functions and press
Start and
Yes
Press the ADJ button on the component board of the HMT330/PTU300 transmitter to open the adjustment mode. The indicator LED on the component board starts to flash.
After this step, the adjustment is carried out using the MI70 indicator.
Press
OK to start the adjustment.
Select RH or T and press
Select.
If you selected RH, check the environment settings if necessary. If checking the environment settings is not necessary, press
No.
Complete the adjustment using 1 of the following alternative methods: