ACAL<cr>

Use the ACAL command to calibrate the currently active analog output mode. You must use this command if you change the analog output mode using the AMODE command.

After you give this command, the analog output of the module is set to the low limit of the selected output (voltage or current). Measure the output and enter the measured value. After entering the low limit, you must measure and enter the high limit also.

Example (analog output in 4 ... 20 mA mode):
>acal
I1 (mA)  ? 3.981
I2 (mA)  ? 20.002
  2.00985160E+03 a0
  1.12789450E+03 a1
  0.00000000E+00 a2
OK
>save
Saving settings...done
>
Example (analog output in 0 ... 1 V mode):
>acal
V1 (V)  ? 0
V2 (V)  ? 0.998
  5.45999950E+02 a0
  1.64128260E+03 a1
  0.00000000E+00 a2
OK
>save
Saving settings...done
>