Filter Types - HMM100

HMM100 User Guide

Document code
M211024EN
Revision
E
Language
English
Product
HMM100
Document type
User guide

All filters available for HMM100 are 12 mm in diameter, and can be used in both the plastic and stainless steel probe. Note the operational temperature range of each filter. For order codes, see Accessories.

Table 1. Filter Types and Properties
Filter Type Diameter Pore Size Temperature Range
Plastic grid (fastest response time) 12 mm - -20 ... +80 °C (-4 ... +176 °F)
Membrane 12 mm 0.2 µm -20 ...  +80 °C (-4 ... +176 °F)
Porous PTFE (general purpose) 12 mm 8 µm -70 ... +180 °C (-94 ... +356 °F)
Stainless steel sintered 12 mm 38 µm -70 ... +180 °C (-94 ... +356 °F)
PPS plastic grid with stainless steel net 12 mm 15 µm -70 ... +180 °C (-94 ... +356 °F)