To view the list of alarms, press Alarms. To view further information of a specific alarm, select an alarm on the screen.
The Device and Site fields show which sensor and site the alarms concern.
Error message | Probable cause | Solution |
---|---|---|
Missing data |
The display is unable to receive a valid Generic Mode Message from the external sending unit. There is a timeout while waiting for data from the external sending unit. The factory default timeout value is 35 seconds. |
Check the external sending unit and connection to it. |
Missing data PAGE_n data N = 1 ... 16 |
The display is unable to receive
valid Generic Mode Messages of the page “ There is a timeout while waiting for
the page “ |
Check the external sending unit and connection to it. |
Any text in the information message field at the bottom of the data pages | The content of the message is defined in the information message part of the Generic Data Message. | |
General parsing error. xxx xxx = More detailed error text. | The display is unable to parse the received Generic Mode Message at all. | Check the format of the Generic Mode Message that is being sent. |