Error messages - WID512

WID512 User Guide

Document code
M211658EN
Revision
F
Language
English
Product
WID512
Document type
User guide

To view the list of alarms, press Alarms. To view further information of a specific alarm, select an alarm on the screen.

The Device and Site fields show which sensor and site the alarms concern.

Table 1. Error messages
Error message Probable cause Solution
Missing data

The display is unable to receive a valid Generic Mode Message from the external sending unit.

There is a timeout while waiting for data from the external sending unit. The factory default timeout value is 35 seconds.

Check the external sending unit and connection to it.
Missing data PAGE_n data N = 1 ... 16

The display is unable to receive valid Generic Mode Messages of the page “n” from the external sending unit.

There is a timeout while waiting for the page “n” data from the external sending unit. The factory default timeout value is 35 seconds.

Check the external sending unit and connection to it.
Any text in the information message field at the bottom of the data pages The content of the message is defined in the information message part of the Generic Data Message.
General parsing error. xxx xxx = More detailed error text. The display is unable to parse the received Generic Mode Message at all. Check the format of the Generic Mode Message that is being sent.