RFL100 Data Logger Modbus registers - AP10 - RFL100 - Jade Smart Cloud

Modbus Data Integration with Jade Smart Cloud and VaiNet Technical Note

Product status
Document code
M213335EN-A
ft:locale
en-US
Product
AP10
RFL100
Jade Smart Cloud
Document type
Technical note
Floating point measurement data registers (read-only)
Register number Address Description Data format Unit
1 0x0000 Relative humidity 32-bit float %RH
3 0x0002 Temperature 32-bit float °C
5 0x0004

Temperature (probe 2)

Available only with probe splitter.

32-bit float °C
9 0x0008 CO2 (%) 32-bit float %
Integer measurement data registers (read-only)
Register number Address Description Data format Scale factor Offset Unit
257 0x0100 Relative humidity 16-bit signed integer 10 0 %RH
258 0x0101 Temperature 16-bit signed integer 10 0 °C
259 0x0102

Temperature (probe 2)

Available only with probe splitter.
16-bit signed integer 10 0 °C
261 0x0104 CO2 (%) 16-bit signed integer 1000 0 %
RFL100 Modbus status data registers (read-only)
Register number Address Description Data format Bitmask / notes
513 0x0200 Error code 16-bit signed integer

0x0000: Status OK

0x0001: Battery low

0x0002: Probe error

0x0004: Probe communication error
514 0x0201 Signal strength 16-bit signed integer Range: 0–4
515 0x0202 Latest measurement time (year) 16-bit signed integer -
516 0x0203 Latest measurement time (month) 16-bit signed integer -
517 0x0204 Latest measurement time (day) 16-bit signed integer -
518 0x0205 Latest measurement time (hour) 16-bit signed integer -
519 0x0206 Latest measurement time (minute) 16-bit signed integer -
520 0x0207 Latest measurement time (second) 16-bit signed integer -
521 0x0208 Available measurement parameters 16-bit signed integer

0x0000: %RH

0x0001: T

0x0002: T (probe 2)

0x0004: CO2 (ppm)

0x0004: CO2 (%)
RFL100 Modbus identification registers (read-only)
Register number Address Description Data format Example contents
7425 0x1D00 Vendor name 8-character ASCII string Constant text: “Vaisala”
7433 0x1D08 Product code 8-character ASCII string “RFL100”
7441 0x1D10 Serial number 8-character ASCII string “S1440571”
7449 0x1D18 Software version 8-character ASCII string “1.2.0”