Refractometer specifications - PR53 - PR53S

PR53S Installation Guide

Product status
Document code
M213333EN-A
ft:locale
en-US
Product
PR53
PR53S
Document type
User guide

Mechanical specifications

PR53S mechanical specifications
Property Specification
Wetted parts
Prism and sapphire plate Sapphire monocrystalline, 99.996 % Al2O3 1
Flow cell Ultra pure PTFE 1
Prism gasket Modified PTFE 1
Process gasket 20.2 × 3 mm (0.8 × 0.1 in) Kalrez W240UP 1
Non-wetted parts
Housing PVDF

Electrical specifications

PR53S inputs and outputs
Property Specification
Supply
Operating voltage 24 V DC nominal (9–30 V DC)
Power consumption Less than 1 W
Protection class 3, PELV
Overvoltage category I
Outputs
Output parameters Refractive index, temperature, concentration, quality factor
Analog outputs
mA

Sourcing, isolated, NAMUR NE 43, configurable

mA range 3.8–20.5 mA
Maximum load 600 Ω
Accuracy of analog outputs at +20 °C ±0.1 % of full scale (±0.0002 RI)
Supported protocol HART 7
Output parameters Concentration
Digital outputs
Digital output RS-485, non-isolated
Maximum cable run 300 m (approx. 1000 ft) (digital)
Supported protocol Modbus RTU
Connectors
External connectors

1 × M12 8-pin A-coded male

With an adapter 1 × M12 4-pin A-coded male

An adapter cable is needed to connect to USB2 and Indigo80. USB2 adapter is used with Insight software, see vaisala.com/insight.

Environmental specifications

PR53S operating environment
Property Specification
Process parameters
Process temperature

–20 … +85 °C (-4 … +185 °F)

Pressure 10 bar at 20 °C, 4.5 bar at 130 °C (145 psi at 68 °F, 652 psi at 266 °F)
Operating environment

Storage temperature

−40 … +65 °C (−40 … +149 °F)

Operating temperature

−20 … +45 °C (−4 … +113 °F)

Maximum operating altitude

2000 m (approx. 6500 ft)

Operating humidity

0–100 %RH

Storage humidity

0–100 %RH, non‑condensing

IP rating

IP66: Dust-tight. Protected from powerful water jets from any direction.

IP67: Dust-tight. Protected from the effects of temporary immersion in water under standardized conditions of pressure and time.

1 Manufacturer's declaration included.