Mechanical specifications
| Property | Specification |
|---|---|
| Wetted parts | |
| Prism and sapphire plate | Sapphire monocrystalline, 99.996 % Al2O3 1 |
| Flow cell | Ultra pure PTFE 1 |
| Prism gasket | Modified PTFE 1 |
| Process gasket 20.2 × 3 mm (0.8 × 0.1 in) | Kalrez W240UP 1 |
| Non-wetted parts | |
| Housing | PVDF |
Electrical specifications
| Property | Specification |
|---|---|
| Supply | |
| Operating voltage | 24 V DC nominal (9–30 V DC) |
| Power consumption | Less than 1 W |
| Protection class | 3, PELV |
| Overvoltage category | I |
| Outputs | |
| Output parameters | Refractive index, temperature, concentration, quality factor |
| Analog outputs | |
| mA |
Sourcing, isolated, NAMUR NE 43, configurable |
| mA range | 3.8–20.5 mA |
| Maximum load | 600 Ω |
| Accuracy of analog outputs at +20 °C | ±0.1 % of full scale (±0.0002 RI) |
| Supported protocol | HART 7 |
| Output parameters | Concentration |
| Digital outputs | |
| Digital output | RS-485, non-isolated |
| Maximum cable run | 300 m (approx. 1000 ft) (digital) |
| Supported protocol | Modbus RTU |
| Connectors | |
| External connectors |
1 × M12 8-pin A-coded male With an adapter 1 × M12 4-pin A-coded male An adapter cable is needed to connect to USB2 and Indigo80. USB2 adapter is used with Insight software, see vaisala.com/insight. |
Environmental specifications
| Property | Specification |
|---|---|
| Process parameters | |
| Process temperature |
–20 … +85 °C (-4 … +185 °F) |
| Pressure | 10 bar at 20 °C, 4.5 bar at 130 °C (145 psi at 68 °F, 652 psi at 266 °F) |
| Operating environment | |
|
Storage temperature |
−40 … +65 °C (−40 … +149 °F) |
|
Operating temperature |
−20 … +45 °C (−4 … +113 °F) |
|
Maximum operating altitude |
2000 m (approx. 6500 ft) |
|
Operating humidity |
0–100 %RH |
|
Storage humidity |
0–100 %RH, non‑condensing |
|
IP rating |