| Change the laser bandwidth factor only when requested to do so by Vaisala technical support. The laser bandwidth factor affects the estimated laser spectrum, which is used in transmission modeling during the water vapor retrieval. When the laser bandwidth factor is not properly set, the water vapor mixing ratio profiles can show dry or wet bias. |
| For optimal performance, Vaisala recommends setting the laser bandwidth factor to match the value in the Test and quality report. The laser bandwidth factor affects the estimated laser spectrum, which is used in transmission modeling during the water vapor retrieval. When the laser bandwidth factor is not properly set, the water vapor mixing ratio profiles can show dry or wet bias. |
You can change the laser bandwidth factor for one measurement unit or both measurement units. The measurement units can have different laser bandwidth factors.
Check that the instrument uses firmware version 2.5, or newer. Update the firmware if needed.