Configuring DA10 laser bandwidth factors - DA10

Configuring DA10 Laser Bandwidth Factor Technical Note

Product status
Document code
M213286EN-C
ft:locale
en-US
Product
DA10
Document type
Technical note
Change the laser bandwidth factor only when requested to do so by Vaisala technical support. The laser bandwidth factor affects the estimated laser spectrum, which is used in transmission modeling during the water vapor retrieval. When the laser bandwidth factor is not properly set, the water vapor mixing ratio profiles can show dry or wet bias.
For optimal performance, Vaisala recommends setting the laser bandwidth factor to match the value in the Test and quality report. The laser bandwidth factor affects the estimated laser spectrum, which is used in transmission modeling during the water vapor retrieval. When the laser bandwidth factor is not properly set, the water vapor mixing ratio profiles can show dry or wet bias.

You can change the laser bandwidth factor for one measurement unit or both measurement units. The measurement units can have different laser bandwidth factors.

Check that the instrument uses firmware version 2.5, or newer. Update the firmware if needed.

  1. Open a connection to DA10.
  2. After logging in, select the DAC measurement unit to modify.

    In the following example, the unit with serial number U1064807 is selected for modification.

  3. Change Laser bandwidth factor to the desired value and select Save.

    The following window opens.

  4. Repeat the steps for the other measurement unit, if needed.