Pump temperature (TP) - RS41-SG - RS41-SGP

Ozone Sounding with DigiCORA software and RS41 User Guide

Document code
M213165EN
Revision
A
ft:locale
en-US
Product
RS41-SG
RS41-SGP
Document type
User guide

DigiCORA applies a pressure-based correction to the measured pump temperature according to equation E-3-15 of GAW Report 268.

T P = T P M + 3.90 0.80 * L o g 10 ( P a i r ) a t P A i r > 3 hPa