Support for Jade Smart Cloud monitoring system and remote firmware update.

Release date: December 2025

New features and improvements

  • Supports factory pre-configuration for connecting to the Jade Smart Cloud monitoring system.
  • Remote firmware update capability from Jade Smart Cloud. This feature can only be used by Vaisala.

Fixes

  • CDF-3008: Fixed an issue where similar probes could be swapped without the new probe indication appearing on the display during probe detection.
  • CDF-3280: Fixed an issue where strings Inactive and Disabled were shown shortened on the display.
  • CDF-3433: Fixed an issue where slow network speed could prevent data from being backfilled to the monitoring system.

Compatibility

  • Compatible with the following monitoring solutions:
    • viewLinc Enterprise Server version 5.2.1 and later. Version 5.2.2 or later required for use with Analog Input Probe ANP115.
    • Jade Smart Cloud.
    At the time of release of this firmware version, Jade Smart Cloud has the following limitations when VDL200 is used with Analog Input Probe ANP115:
    • Boolean mode is not supported.
    • Door switch mode is not supported.
    • ANP115 door switch model is not supported.
  • Compatible with Insight PC software version 1.3.0 and later.

Upgradeability

  • Can be installed on any VDL200 data logger that has an earlier firmware version.