- 1
- Power switch
- 2
- Probe splitter
- 3
- Latch of the battery cover
- 4
- Instruction label with short version of this setup instruction
- 5
- Probe connectors of the probe splitter
- 6
- Temperature probes
- 7
- Numbering labels for TMP115 probes (included in the probe mounting accessories)
|
| NOTICE
You must follow the procedure below to set up RFL100 with the probe splitter. The procedure includes steps to label the probes to make it easy to identify the channel assignment later. You can also use the info mode of RFL100 to check which probe serial number is connected to which channel. For more information on the info mode, see RFL100 Data Logger User Guide (M211861EN). |
|
Once the data logger has been set up for 2 temperature probes using the probe splitter, it requires a factory reset before it can be set up for one-probe mode or operation with a CO2 probe. For the factory reset procedure, see RFL100 Data Logger User Guide (M211861EN). |