Vaisala RFL100 data logger is a wireless, battery powered data logger. It supports several types of Vaisala probes for measurement of humidity, temperature, and carbon dioxide (CO2). RFL100 operates as a data collection point in a Vaisala monitoring system.
The wireless connection of RFL100 requires a Vaisala AP10 access point. A single AP10 can connect up to 32 RFL100 data loggers. In a typical indoor space, install the AP10 within 100 meters (approx. 330 ft) of the RFL100. In an open space with line-of-sight and no interfering structures, the range can be over 500 m (1640 ft).
RFL100 is optimized for low power operation. It reads the probe once a minute, and transmits measurement data to the access point every 4 minutes. Because the radio link is not continuous, remote management actions and system joining status may take some time to be updated on the display of the data logger.
|
Before you start installing RFL100 data loggers, install one or more AP10 access points within range of RFL100. This allows RFL100 to immediately connect to your monitoring system. |