| Err 100 |
User parameter bank checksum failure. |
Power cycle the RFL100. If the error persists, contact
Vaisala. |
| Err 101 |
Factory parameter bank checksum failure. |
| Err 102 |
Real-time clock has lost accurate time. |
Restore the wireless connection to an AP10 access point. RFL100 will
synchronize its clock with the time from the access point. |
| Err 103 |
Main battery voltage is critically low. When this error appears, RFL100 will soon
stop radio communication to conserve energy. It will continue to record measurement
data in the local memory as long as possible. |
Replace the main batteries of RFL100. |
| External power supply is connected but the power switch is in the Off position. |
Move the power switch to the On position. Note that it will
take some time for the error to clear if RFL100 remains continuously powered. |
| Err 104 |
Incorrect factory configuration parameters. |
Power cycle the RFL100. If the error persists, contact
Vaisala. |
| Err 105 |
Real-time clock hardware error. |
| Err 106 |
RFL100 is operating in CO2 mode without external power. |
Connect an external power supply to the service port of RFL100 to avoid draining
the batteries. |
| Err 110 |
If using RFL100 with 1 probe:
- No probe connected at
startup
- Incompatible probe detected
- Trying to switch from 2-probe
mode to 1-probe mode without performing a factory reset
|
- Verify that the probe is a
compatible model.
- Power cycle the RFL100 to
re-detect the probe.
- If the error persists, it is
likely that the RFL100 has been set up in 2-probe mode. Perform the factory reset
procedure to allow 1-probe use again, then power cycle the RFL100 to re-detect the
probe.
|
If using 2 temperature probes with the probe splitter:
- No probes connected at
startup
- Two probes connected at same time
without following proper setup or probe swap procedure
- RFL100 is in 2-probe mode but 1
probe is missing
- Unsupported probe
combination
- At least 1 incompatible probe
detected
|
- Verify that probe splitter and 2
supported temperature-only probes are connected to the RFL100.
- If you reconnected any probes,
power cycle the RFL100 to re-detect the probe(s).
- If the error persists, perform
the 2-probe setup again.
|
| Err 200 |
Real-time clock battery voltage is low. |
Replace the clock battery. |
| Err 202 |
Communication failure with at least 1 probe that was detected at
startup. |
- Check that the probe(s) are connected properly.
- If you reconnected any probes,
wait 1 minute for the display to update and verify that the error is gone.
- If the error persists, power
cycle the RFL100 to re-detect the probe(s).
|
| Err 203 |
Probe error. Can be caused by probe damage or a wet humidity
sensor. |
Inspect the probe and replace it if necessary. If the error has been
caused by a wet humidity sensor, wait for it to dry out. |
| Err 204 |
Real-time clock temperature compensation problem. |
If error code Err 200 is also active, replace
the clock battery. If the error persists, contact Vaisala. |