| Issue | Remedy |
|---|---|
| When you change the sensor or calibration settings for DSC211 in the Web UI, the sensor restarts to take the new settings into use. This may show as an error state during the reboot. | After the next sensor poll (maximum 10 s), the sensor state returns to OK. |
| Serial console bypass does not work with DSC211 and DST11, making it impossible to establish a terminal connection to the sensors using the serial console in the Web UI. |
Make note of the following station parameters. Then change them as indicated below, and restart the script.
When done, restore the original settings and restart the script. |
| DSC211 is unable to obtain measurements from DST111 (surface temperature, air temperature, and relative humidity). |
Using the serial console bypass,
set the DSC211 CONF 4 [Slave A] parameter
to DST111 address ( When the ID is set, the observations from DST111 are included in the DSC211 data messages. DSC211 can also use the data in autocalibration and in frost detection. |
| In the Web UI, power measurement does not show actual values. | If you need to know the power consumption, use PSU501 measurements (PSU OUTPUT CURRENT and PSU OUTPUT VOLTAGE) in the Web UI to calculate it. |