You can configure the analog output over Modbus RTU or by using Vaisala Insight PC software.
Configure the output for wind speed and wind direction separately.
- Create a connection to the device with Insight PC software.
- Select the settings icon () and Configure device.
-
Select Analog output, wind speed and configure the following settings.
- Output mode
- Fixed to 4–20 mA.
- Output parameter
- Select sample, instant, or average as the measurement value.
- Scale low end
- The wind speed value 4.0 mA output represents.
- Scale high end
- The wind speed value 20.0 mA output represents.
- Error output level
- Current indicating device failure.
- Output level
- Read-only. The actual (calculated) electrical mA level of the output.
- Test output level
- Set the test output level.
- Select Save.
-
Select Analog output, wind direction and configure the following settings.
- Output mode
- Fixed to 4–20 mA.
- Output parameter
- Select sample, instant, or average as the measurement value.
- Error output level
- Current indicating device failure.
- Output level
- Read-only. The actual (calculated) electrical mA level of the output.
- Test output level
- Set the test output level.