Refractometer specifications - PR53 - PR53SD - SDI5

PR53SD and SDI5 Safe-Drive Installation and Safety Guide

Document code
M212953EN
Revision
D
Language
English
Product
PR53
PR53SD
SDI5
Document type
User guide

Mechanical specifications

PR53SD mechanical specifications
Property Specification
Wetted parts
Sensor head EN 1.4462 1
Prism Sapphire monocrystalline, 99.996 % Al2O3 2
Process gasket Co-Cr-Ni Alloy (AMS 5876) lined PTFE 2
Prism gasket Modified PTFE 2
SD flange EN 1.4462 1
Wash nozzle EN 1.4462 2
Non-wetted parts
Housing EN 1.4404
Screws, TX20 torque 2.0 Nm EN 1.4404 (AISI 316L)
Stud bolts, M12 torque 75 Nm, M10 torque 40 Nm EN 1.4435 (AISI 316L), grade 8.8
Flanges (3 pcs)

EN 1.4462 (AISI 2205)

ASME B16.5, DIN 2543

Cable 2×2×0.5 mm2 (AWG 21), PUR jacket, grey 10 m multistrand, with ferrules

Flame-retardant acc. to IEC 60332-1-2, FT1, VW1

Weight

Retractor and wash connection 15 kg (33.07 lb)

Refractometer 5.4 kg (11.9 lb)

Electrical specifications

PR53 series inputs and outputs
Property Specification
Supply
Operating voltage 24 V DC nominal (9–30 V DC)
Power consumption Less than 1 W
Protection class 3, PELV
Outputs
Output parameters RI, temperature, concentration, quality factor
Analog outputs
mA

Sourcing, isolated, NAMUR NE 43, configurable

mA range 3.8–20.5 mA
Loop impedance Max. 600 Ω
Accuracy of analog outputs at +20 °C ±0.1 % of full scale (±0.00002 RI)
Digital outputs
Digital output RS-485, non-isolated
Maximum cable run 300 m (approx. 1000 ft) (digital)
Supported protocol Modbus RTU
Connectors
External connectors

1 × M12 M 4 pins, A-coded 3

2 × M16×1.5 cable gland, Cable D 5–10 mm / Adapter for conduit entry M16×1.5 4/ NPT ½"

Environmental specifications

PR53SD operating environment
Property Specification
Process parameters
Process temperature –40 … +170 °C (–40 … +338 °F)
Design temperature +180 °C (+356 °F) 5
Design pressure / maximum operating pressure 35 bar (507.6 psi)
Maximum retraction pressure 35 bar (507.6 psi)
Operating environment

Storage temperature

−40 … +65 °C (−40 … +149 °F)

Operating temperature

−40 … +60 °C (−40 … +140 °F)

Maximum operating altitude

2000 m (approx. 6500 ft)

Operating humidity

0–100 %RH

Storage humidity

0–100 %RH, non‑condensing

UL 50E (NEMA) rating

Type 4X: Dust-tight. Protected from corrosion and hose-directed water.

IP rating

IP66: Dust-tight. Protected from powerful water jets from any direction.

IP67: Dust-tight. Protected from the effects of temporary immersion in water under standardized conditions of pressure and time.

1 EN 10204 / 3.1 certificate included.
2 Manufacturer's declaration included.
3 For USB2 adapter and Insight software, see www.vaisala.com/insight.
4 Conduit hub is not compatible with SDI5 Safe-Drive system.
5 Maximum momentary temperature peak.