Before starting, locate the emergency shower.
  • Refractometer
  • 18‑mm wrench
Follow these instructions to install the refractometer into the retractor. Contact Vaisala technical support for further assistance if needed.
Scan the QR code to watch a video of the installation and removal of the refractometer: https://youtu.be/UVMlfPiz3-4?si=1OQtPKXKT741jnO_

For written instructions on the removal, see PR53 Series User Guide (M212898EN).

  1. Familiarize yourself with the safety instructions of the installation site.
  2. Make sure the isolation valve is closed.
  3. Place the refractometer into the retractor. Make sure the cable glands on the refractometer are in a downward position.
  4. Rotate the hand wheel until it stops.
    CAUTION Do not force the wheel.
  5. Open the isolation valve.
  6. Rotate the hand wheel until it stops.
  7. Install the nuts and tighten them using a 18-mm wrench.
  8. To connect refractometer wiring, see Connecting the refractometer wiring.