Probe options - VDL200

VDL200 User Guide

Document code
M212938EN
Revision
A
Language
English
Product
VDL200
Document type
User guide
Probe Description 1 Installation notes

HMP110 and HMP110T

Humidity and temperature probe for measurement in demanding conditions. Robust stainless steel construction. Temperature-only version HMP110T available.

Plastic grid filter provides the fastest response time. For added protection, select the membrane filter, the PTFE filter, or the stainless steel sintered filter.

Measurement temperature range −40 … +80 °C (−40 … +176 °F).

Suitable for measurement inside chambers, incubators, fridges, and freezers.

Versatile mounting options using accessories.

Must be connected to VDL200 using an M8/M8 probe cable (for example, CBL211293-3MSP).

HMP115 and HMP115T

Humidity and temperature probe for general purpose measurement. Temperature-only version HMP115T available.

Plastic grid filter provides the fastest response time. For added protection, select the membrane filter or the PTFE filter.

Measurement temperature range −40 … +60 °C (−40 … +140 °F).

Ideal choice for ambient measurement.

Can be connected to VDL200 directly or using an M8/M8 probe cable (for example, CBL211293-3MSP).

TMP115

Temperature probe for measurement in a wide range of conditions.

Available as 50 cm (1 ft 7.7 in) and 3 m (9.8 ft) long versions. Length includes the probe body and sensor tip.

Measurement temperature range −196 … +150 °C (−320 … +302 °F).

Operating temperature range of the probe body is −40 … +60 °C (−40 … +140 °F).

Suitable for measurement inside chambers, fridges, and freezers. Sensor tip withstands immersion in glycol and liquid nitrogen.

Can be connected to VDL200 directly or using an M8/M8 probe cable (for example, CBL211293-3MSP).

Use the thermal dampener block accessory (item code 236310SP) to add thermal mass to the sensor tip.

GMP251 and GMP252

Carbon dioxide probes for %-level measurements (GMP251) and ppm-level measurements (GMP252). Designed for use in demanding applications such as life science incubators.

Measurement temperature range −40 … +60 °C (−40 … +140 °F).

When ordered with the VDL200 data logger, the probe is delivered with an M8-M12 adapter cable (item code CBL211291SP) and the CO2 probe mounting kit (item code ASM214253SP). The kit includes a probe holder that has attachments for the CO2 probe, a Ø 12 mm probe (for example, HMP110) and the sensor tip of the TMP115 probe.

Must be connected to VDL200 using the M8/M12 flat cable (item code CBL211291SP).

1 See probe datasheets in docs.vaisala.com for detailed probe specifications.