VDL200 detects and validates the connected probe(s) only at startup. Because of this, you must restart the data logger to take any probe changes into use. Probe cables can be disconnected and reconnected on the fly, as long as the probes and their connection ports remain the same.

If you change the probes in a way that changes the parameters recorded by the data logger, log in to your monitoring system and check if you need to make changes to way measurement data is linked to locations.

  1. Remove the data logger from the mounting bracket.
  2. Open the battery cover of the data logger.
  3. Push the power button to turn off the data logger.
  4. Disconnect and connect probes as needed.
  5. Push the power button to turn on the data logger.
  6. Close the battery cover of the data logger.
  7. Check the display to verify that the connected probes are successfully detected.
  8. Insert the data logger back in the mounting bracket.