Manage Devices
Channel calibration settings are the reference values used for testing measurement accuracy on Vaisala data loggers
Ensure that the initial calibration values have been downloaded from the device to viewLinc (automatically discovered when device detected by viewLinc). |
- In Sites Manager select the Hosts and Devices tab.
- In the Hosts and Devices tree, select a data logger channel to edit.
- Select .
- In the Edit Channel Properties window, edit the properties, Calibration scale and Calibration offset using information provided by Vaisala or collected from on-site calibration testing.
- If you are changing calibration settings on an HMT140 device, you are prompted to reset the device calibration settings.
- Select OK.