Installation environment - PR53 - PR53APEX - PR53AP

PR53AP and PR53APEX Installation Guide

Product status
Document code
M212932EN
Revision
D
ft:locale
en-US
Product
PR53
PR53APEX
PR53AP
Document type
User guide

Follow these guidelines to ensure the best operating environment for the process refractometer:

  • Install the refractometer indoors or outdoors. In outdoor installations, protect the device from exposure to sunlight and rain.
  • If the pipe is translucent, protect the prism from light. Outside light may disturb the measurements.
  • Use additional cooling according to Process temperature and recommended cooling. Notice that the values in the figure are indicative. Required cooling depends on the installation location and environment.
Process temperature and recommended cooling