Temperature measurement (Pt100 sensor)
1
2 |
Measurement range |
−60 … +85 °C (−76 … +185 °F) |
Uncertainty over temperature range −60 … +85 °C (−76 … +185 °F) |
< ±0.04 °C |
Typical error at 0 °C (+32 °F) 3 |
< ±0.01 °C |
Long‑term stability |
To be defined |
Temperature measurement (Pt1000 sensor)
1
2 |
Measurement range |
−60 … +85 °C (−76 … +185 °F) |
Uncertainty over temperature range −60 … +85 °C (−76 … +185 °F) |
< ±0.35 °C |
Typical error at 0 °C (+32 °F) 3 |
< ±0.2 °C |
Long‑term stability |
To be defined |
Voltage measurement |
Uncertainty over temperature range −40 … +70 °C
(−40 … +158 °F): |
±2.5 V range |
< 0.05 % of reading ±1.85 μV |
±1.25 V range |
< 0.04 % of reading ±2.3 μV |
±620 mV range |
< 0.04 % of reading ±0.9 μV |
±310 mV range |
< 0.04 % of reading ±0.75 μV |
±156 mV range |
< 0.04 % of reading ±0.8 μV |
±78 mV range |
< 0.04 % of reading ±1.25 μV |
±39 mV range |
< 0.04 % of reading ±0.8 μV |
Uncertainty over temperature range −60 … +85 °C
(−76 … +185 °F): |
±2.5 V range |
< 0.05 % of reading ±1.85 μV |
±1.25 V range |
< 0.04 % of reading ±2.3 μV |
±620 mV range |
< 0.04 % of reading ±0.9 μV |
±310 mV range |
< 0.04 % of reading ±0.85 μV |
±156 mV range |
< 0.04 % of reading ±1.2 μV |
±78 mV range |
< 0.05 % of reading ±1.25 μV |
±39 mV range |
< 0.06 % of reading ±0.95 μV |
Common mode range |
+2.5 V / −0.9 V |
Voltage measurement 0–30 V |
|
Uncertainty over temperature range −40 … +70 °C
(−40 … +158 °F): |
To be defined |
Uncertainty over temperature range −60 … +85 °C
(−76 … +185 °F): |
To be defined |