RI calibration - PR53 - PR53AC - PR53AP - PR53GC - PR53GP - PR53M - PR53SD - PR53W

PR53 Series User Guide

Document code
M212898EN
Revision
F
ft:locale
en-US
Product
PR53
PR53AC
PR53AP
PR53GC
PR53GP
PR53M
PR53SD
PR53W
Document type
User guide

In order to perform RI calibration, you need:

  • Indigo520 transmitter or Vaisala Insight software
  • Sample holder
  • Cleaning solution
  • Tissues
  • Nominal reference RI liquids, see below
  • Protective clothing
  • Environment with a good ventilation

Nominal reference RI liquids

RI calibration is performed using a set of nominal reference RI liquids that cover the entire measurement range of the prism. The selection of nominal reference RI liquids varies depending on the type of prism in use, as different types of prisms cover a specific RI measurement range.

  • A = RI 1.320–1.530 nD Sapphire prism
  • B = RI 1.360–1.570 nD Sapphire prism

The prism type can be identified from the configuration code located on the side of the refractometer.

For refractometer models PR53AC, PR53AP, PR53GC, PR53GP, and PR53SD, refer to the 7th digit of the configuration code.

For refractometer models PR53M and PR53W, refer to the 6th digit of the configuration code.

CAUTION Each liquid has their own safety instructions with them. Read the instructions carefully before you start RI calibration.