Upload concentration curves if you need
to restore original concentration curves, add new previously not configured concentration
curves, or take into use updated concentration curves.
Log into MyVaisala.com and go to LM
Calibration tools > Concentration models POLARIS PR53. If you do not have access to the tools, contact your local distributor for
the curve data strings.
Use the search function to find the correct concentration curve.
Select the concentration curve from the list to view its details.
Select Copy
to clipboard to copy the curve data string.
The curve data string is copied to your clipboard.
In Insight, go to > Configure device > Concentration curve upload.
Select a target concentration curve from the Target curve drop down menu. You can either overwrite one of the existing concentration curves or
select Add a new curve. You can add a maximum of 4 concentration curves.
Paste curve data string into the Curve data string field.
Accept the change by selecting Save.
It takes 10–15s to save the changes. The refractometer may restart.
If you
are replacing an active concentration curve, the new curve is automatically taken into
use. See Changing concentration curves for
activating concentration curves.