With PR53 refractometers it is possible to use configurable measurement damping.
Measurement damping reduces noise in measurement and adjusts the response time of the measurement to correspond to the response time of the process and the related control circuits.
There are 3 types of measurement damping:
- Exponential damping
- Linear damping
- Slew rate
The default damping settings for PR53 refractometers are:
- Damping type: exponential
- Damping time: 20 s
- Tolerance time: 5 s
- Slew rate: 10.0.