Measurement status details - PR53 - PR53AC - PR53AP - PR53GC - PR53GP - PR53M - PR53SD - PR53W

PR53 Series User Guide

Document code
M212898EN
Revision
F
ft:locale
en-US
Product
PR53
PR53AC
PR53AP
PR53GC
PR53GP
PR53M
PR53SD
PR53W
Document type
User guide

There are different measurement statuses that you can view in the Diagnostics tab of the Insight software.

  • C measurement status
  • T measurement status
  • Field sample status
Measurement status details
Message C measurement status T measurement status Field sample status
None No errors detected. No errors detected. No errors detected.
Unavailable (S) Concentration measurement is unavailable because no liquid is detected on the prism or the device is in calibration mode. - No field sample has been taken.
Unreliable (R) Refractometer is measuring temperature outside concentration curve measurement range. Temperature compensation might not work properly. - A warning in image detection has occurred during taking the field sample.
Under range (U) Refractometer is measuring a value outside concentration curve measurement range. Refractometer is measuring a value outside concentration curve measurement range. -
Over range (O) Refractometer is measuring a value outside concentration curve measurement range. Refractometer is measuring a value outside concentration curve measurement range. -
Locked (L) Concentration value is locked during prism wash. Temperature value is locked during prism wash. -
Failure (F)

Optical image is not available.

Refractometer needs maintenance, contact helpdesk@vaisala.com.

Temperature sensor has failed.

Refractometer needs maintenance, contact helpdesk@vaisala.com.

Optical image is not available or temperature sensor has failed during taking the field sample.
Not ready (W) - - Field sample is being measured and the results are not yet available.