Exponential damping works for most processes and is the standard choice for slow and continuous processes. The factory setting is always exponential damping.
Select Damping type to switch between different damping algorithms.
In the exponential damping, the damping time is the time it takes for the concentration measurement to reach half of its final value at a step change. For example, if the concentration changes from 50 % to 60 % and damping time is 10 s, it takes 10 s for the Indigo520 to display concentration 55 %. A damping time of 5–15 s works best in most cases, the factory setting is 5 s.
Select Damping time to set the damping time.