Follow these instructions to replace the
2.5" EPDM gasket in PR53 refractometers and flow cells.
Installing 2.5" EPDM gasket to PR53AC
refractometer
Installing 2.5" EPDM gasket to flow cells
-
Familiarize yourself with the safety instructions of the installation site.
-
Make sure the process line is depressurized and empty.
-
Open the clamp or clamps holding the refractometer or flow cell in place.
-
Replace the gasket.
| CAUTION The old gasket can be hard to remove. To avoid damaging the sealing
surface, do not use sharp tools for removing or installing the gaskets. |
-
Reinstall the refractometer with clamps.
-
Keep the material certificates delivered with the spare part for later use.
-
Recycle removed parts according to the recycling instructions in PR53 Series
User Guide (M212898EN).