The reference value needs to be known.
- Go to > Calibrate device.
- Insert your device in the reference environment for the first calibration point.
- Wait for the measurement to stabilize fully. The graph shows readings of the last 60 minutes.
- When the measurement has stabilized, select the Reference value, point 1 text box and enter the temperature of calibration point 1. Select Enter or click outside the field when done.
- Check that the measured value for point 1 is automatically inserted.
-
Check the difference between each reference and measured value. Very large differences
may be due to insufficient stabilization time or unsuitable calibration setup.
If you want to adjust the device, select Activate adjustment and verify the result from the message that appears at the top of the screen..
To exit without taking the adjustment in use, select Close.
- After calibrating your device, update the information in the Calibration information tab.