Problem | Severity level | Cause | Corrective action |
---|---|---|---|
Message Internal humidity too high | Error |
The relative humidity measured at the refractometer's processor card exceeds 60 %RH. The reason may be moisture leaking in through the prism seal or the cover being open. |
Make sure that the prism seal is intact and the cover is closed. If necessary, replace the prism seal. |
Message Internal temperature too high | Error |
The temperature on the sensor
processor card exceeds +65 °C (+150 °F). To read this temperature,
|
Make sure you are using the device in the right temperature. See the ambient and process temperature in the model-specific specifications. |
Indigo520: No devices connected / Disconnected Insight: No connected devices in the start screen |
n/a |
The cables are not properly connected. The refractometer is connected to both Indigo520 and Insight. |
Make sure that cables are properly connected at both ends. If the refractometer is connected to both Indigo520 and Insight, it prioritizes Indigo520 and is not visible to Insight. If you are able to open the refractometer cover, make sure that the green LED is on. This means that the refractometer is powered. |
Relays not working | n/a |
Relays may be in test mode. Make sure using Indigo520 that the test mode is inactive in .For testing the wash function, see PR53 Prism Wash System User Guide (M212808EN). |
|
Analog output signal is not working as expected | n/a |
Check wiring, see the model-specific installation guide. If the analog output signal does not correspond to the concentration display, check output signal configuration, see Diagnostics. For possible correction, configure
the analog output.
A low analog output signal can also be caused by high resistance in the external current loop, see Refractometer connections. A noisy signal can be damped, see
|
Problem | Severity level | Cause | Corrective action | ||
---|---|---|---|---|---|
Message No optical image | Critical |
To see the optical image, you need to view the Diagnostics menu in Indigo520 or Insight. There are several possible causes:
|
|||
Message Blank image corrupted | Critical | Blank image is missing or corrupted. | Create a new blank image. See Blank image. Contact helpdesk@vaisala.com. | ||
Message External light level too high | Error | The measurement is not possible because too much outside light reaches the camera. | Identify the light source (for example sun shining into an open tank or a translucent pipe) and block the light from getting to the prism at the sensor tip. | ||
Message Prism coating detected | Error | The optical surface of the prism is coated by the process medium or impurities in the process medium. |
Perform prism wash if available, see PR53 Prism Wash System User Guide (M212808EN). If prism wash is not available, remove the refractometer from process line and clean prism manually. If the problem is recurrent, consider improving the flow conditions, see model-specific installation guide. If prism wash is available, adjust the wash parameters, see PR53 Prism Wash System User Guide (M212808EN). |
||
Message Temperature measurement error | Error | A temperature element is faulty. |
Contact your distributor or helpdesk@vaisala.com, the temperature element may need to be replaced. Note that a difference to some other process temperature measurement is not a fault. PR53 measures the true temperature of the prism surface. |
||
Message Invalid calculation parameters | Error | The A parameters are incorrect. | Perform RI adjustment or contact helpdesk@vaisala.com. | ||
Message C measurement out of concentration curve range | Warning | Concentration measurement is outside of the range defined in the concentration curve. If the measurement concentration is other than the valid range for the used curve, the measurement result may be unreliable. |
|
||
Message T measurement out of concentration curve range | Warning | Temperature measurement is outside of the range defined in the concentration curve. If the measurement temperature is other than the valid range for the used curve, the measurement result may be unreliable. |
|
||
Message External light level high | Warning | Some light from the outside reaches the sensor and may disturb the measurement. | Identify the light source (for example sun shining into an open tank or a translucent pipe) and block the light from getting to the prism at the sensor tip. | ||
Message Image quality low | Warning | The most likely cause for this message is scaling on the prism. There still is an optical image available, but the measurement quality may not be optimal. | Clean the prism. | ||
Message No liquid detected | Warning | The operation of the equipment is OK but there is no process liquid on the prism. | |||
Message Calibration mode active | Warning | The refractometer can be set to calibration mode in the context of RI calibration check. When calibration mode is active, the refractometer will not calculate the concentration. | Make sure that the calibration mode is off in all used user interfaces. | ||
Unexpected concentration drift | n/a |
For drift upward, suspect prism coating, see PR53 Prism Wash System User Guide (M212808EN). Otherwise make sure calibration is correct, see RI calibration, as well as RI adjustment, see RI adjustment. The latter can only be done with Insight. A good way to observe the possible concentration drift is to check the measurement log from Indigo520, if available. In addition to RI and concentration, this log includes the QF value, the changes of which can indicate prism coating. See Indigo500 User Guide (M212287EN). |