WXT534 specifications - DA10

DA10 User Guide

Document code
M212895EN
Revision
D
ft:locale
en-US
Product
DA10
Document type
User guide
Table 1. WXT530 barometric pressure measurement performance
Property Description/Value
Observation range 500–1100 hPa
Accuracy (for sensor element) at 600–1100 hPa

±0.5 hPa at 0 … +30 °C (+32 … +86 °F)

±1 hPa at −52 … +60 °C (−60 … +140 °F)

Output resolution 0.1 hPa / 10 Pa / 0.001 bar / 0.1 mmHg / 0.01 inHg
Units available hPa, Pa, bar, mmHg, inHg
Table 2. WXT530 air temperature measurement performance
Property Description/Value
Observation range −52 … +60 °C (−60 … +140 °F)
Accuracy (for sensor element) at +20 °C (+68 °F) 1 ±0.3 °C (±0.54 °F)
Output resolution 0.1 °C (0.1 °F)
Units available °C, °F
Table 3. WXT530 relative humidity measurement performance
Property Description/Value
Observation range 0–100 %RH
Accuracy (for sensor element) 2

±3 %RH at 0–90 %RH

±5 %RH at 90–100 %RH

Output resolution 0.1 %RH
PTU measuring interval 1–3600 s (= 60 min), at 1 s steps
Table 4. WXT530 inputs and outputs
Property Description/Value
Operating voltage 6–24 V DC (−10 … +30 %)
Average power consumption

Minimum: 0.1 mA at 12 V DC (SDI‑12 standby)

Typical: 3.5 mA at 12 V DC (typical measuring intervals) 3

Maximum: 15 mA at 6 V DC (constant measurement of all parameters)

Heating voltage

DC, AC, or full-wave rectified AC

12–24 V DC (−10 … +30 %)

12–17 V ACrms (−10 … +30 %)

Typical heating current

12 V DC: 800 mA, 24 V DC: 400 mA

12 V ACrms: 1.1 A

17 V ACrms: 800 mA

Digital outputs SDI‑12, RS‑232, RS‑485, RS‑422
Communication protocols

SDI‑12 v1.3, Modbus RTU, ASCII automatic and polled, NMEA 0183 v3.0 with query option

Self-diagnostic Separate supervisor message, unit/status fields to validate measurement stability
Startup Automatic, < 5 seconds from power on to the first valid output
Table 5. WXT530 operating environment
Property Description/Value

Operating environment

Outdoor use

Operating temperature

−52 … +60 °C (−60 … +140 °F)

Storage temperature

−60 … +70 °C (−76 … +158 °F)

Operating humidity

0–100 %RH

Operating pressure

500–1100 hPa
Wind 4 0–60 m/s (0–134 mph)

IP rating

Without mounting kit: IP65: Dust-tight. Protected from water jets from any direction.

With mounting kit: IP66: Dust-tight. Protected from powerful water jets from any direction.

Table 6. WXT530 compliance
Property Description/Value

EU directives and regulations

EMC Directive (2014/30/EU)

RoHS Directive (2011/65/EU) as amended by 2015/863

Electromagnetic compatibility (EMC)

EN 61326‑1, industrial environment

CISPR 32 / EN 55032, Class B

Environmental

IEC 60068‑2‑1, 2, 6, 14, 30, 31, 78

IEC 60529, VDA 621‑415

Maritime

IEC 60945 (Exposed)

DNV GL Type Examination Certificate No. TAA00000VF

Compliance marks

CE, RCM, RoHS, China RoHS, UKCA
Table 7. WXT530 mechanical specifications
Property Description/Value
Weight
0.7 kg (1.54 lb)
Materials
Radiation shield, top, and bottom parts Polycarbonate +20 % fiberglass
Precipitation sensor plate Stainless steel (AISI 316)
Figure 1. WXT535 and WXT534 dimensions
1 A naturally aspirated radiation shield is applied which can affect readings in calm wind.
2 A naturally aspirated radiation shield is applied which can affect readings in calm wind.
3 Wind 10‑second average with 2‑minute interval at 4 Hz sampling rate, RS‑232 19200 bps with jumper wires, PTU 10‑second interval, Pt1000, level, tipping bucket, and solar radiation 5‑second interval.
4 Due to the measurement frequency used in the sonic transducers, RF interference in the 200–400 kHz range can disturb wind measurement.