DA10 maintenance - DA10

DA10 User Guide

Document code
M212895EN
Revision
D
ft:locale
en-US
Product
DA10
Document type
User guide

The instrument's self-diagnostic system runs in the background and generates warnings and alarms according to the issues discovered. Malfunctions are reported in netCDF and status messages.

DA10 requires the following periodic maintenance.

Table 1. DA10 maintenance
Task Frequency
Clean the windows. Every 6 months
Perform a general check. Annually
Check the door gaskets. Annually
Clean and maintain painted surfaces. Annually
Replace the PTU module in WXT534. Every 2 years
Replace the DMU801-DA10 lithium battery. Every 15 years
WARNING Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. Return the instrument to a Vaisala office or authorized depot for service and repair to make sure that safety features are maintained.
WARNING Verify outdoor installation grounding periodically to minimize shock hazard.
WARNING Do not perform periodic maintenance on the system alone. Do not reach into parts and assemblies that are AC (mains) powered and live except in the presence of someone who can provide first aid.