Figure 1. Weight of main units and lifting
handles and locations
1
Lifting locations
2
Lifting handles
DA10 is delivered with its main cabling
connected inside the radiation shield.
WARNING For eye safety reasons, DA10 must be installed vertically.
Lift the radiation shield on the foundation screws.
Use a spirit level to check that the
instrument pedestal is leveled horizontally.
WARNING When the measurement
units are in place do not use the lifting hooks for lifting the instrument. The lifting
hooks cannot support the full weight of the instrument. The lifting hooks are for
lifting the radiation shield when empty.
1
Lifting hooks (2 pcs)
Attach the washers and nuts and tighten the nuts.
Open the back door and remove the bottom cover to make placing the washers and
tightening the nuts easier.
1
Hex nut M10 DIN934 A4 (4 pcs)
2
Flat washer M10 DIN125 A4 (4 pcs)
3
Foundation screw (4 pcs)
4
Instrument pedestal
5
Foundation
There are 4 extra nuts and washers in the accessories package. They are not needed
during installation.
Mount the far-range measurement unit inside the radiation shield.