These are general instructions for the
refractometer mounting. Plan the process‑specific installation according to your requirements.
Contact Vaisala helpdesk for further assistance if needed.
-
Familiarize yourself with the safety instructions of the installation site.
-
Make sure the process line is depressurized and empty.
-
Install the mounting equipment into to the process line.
-
If you are mounting with a flange, weld the flange half into the process
pipeline.
-
If you are mounting with a clamp, weld the ferrule into the process pipeline.
-
Make sure the wash plug is firmly in place so that there is no leakage. If your
refractometer will have a wash system, install the wash system non-return valve before
mounting the refractometer.
-
Mount the refractometer. Make sure the cable glands on the refractometer are in a
downward position.
| Be careful not to damage the wash nozzle at the end of the probe when inserting it
into the pipe. |
-
Install the gasket and tighten the flange bolts. Tighten according to the strength
class of the acquired bolts.
-
Install the O-ring and tighten clamp bolts to 44 Nm with a 17‑mm wrench.
-
Make sure the refractometer is firmly in place and that there is no excessive
vibration. Support the piping if needed.