Installation environment - PR53 - PR53AC

PR53AC Installation Guide

Document code
M212866EN
Revision
C
Language
English
Product
PR53
PR53AC
Document type
User guide

Follow these guidelines to ensure the best operating environment for the process refractometer:

  • You can install the refractometer indoors or outdoors. In outdoor installations, protect the device from exposure to sunlight and rain.
  • If the pipe is translucent, protect the prism from light. Ambient light may disturb the measurements.
  • Use additional cooling according to Process temperature and recommended cooling. Notice that the values in the figure are indicative. Required cooling depends on the installation location and environment.
Process temperature and recommended cooling