Refractometer specifications - PR53 - PR53AC

PR53AC Installation Guide

Document code
M212866EN
Revision
C
Language
English
Product
PR53
PR53AC
Document type
User guide

Mechanical specifications

PR53AC mechanical specifications
Property Specification
Wetted parts
Sensor head EN 1.4435 BN2 (AISI 316L) 1
Surface roughness Ra 0.8 μm
Prism Sapphire monocrystalline, 99.996 % Al2O3 2
Prism gasket Modified PTFE 3
Sanitary 2.5" gasket

EPDM 2

Type N gasket EPDM 2
Welding ferrule EN 1.4435 (AISI 316L) 1 4

ASME BPE-2019 (DIN 32676-C)

Non-wetted parts
Housing EN 1.4404 (AISI 316L)
Screws TX20, torque 2.0 Nm EN 1.4404 (AISI 316L)
Cable gland

EN 1.4305 (AISI 303)

HUMMEL 1.693.1600.50

Dummy plug

EN 1.4305 (AISI 303)

AGRO 8717.96.08.70

Conduit hub

EN 1.4404 (AISI 316L)

Vaisala, DRW257718, M16×1.5 / NPT ½ in

M12 connector

Gland, EN 1.4305 (AISI 303)

Contacts, CuZn with Ni/Au plating

Phoenix Contact, 1405233, M12/4(M), A, 4×0.34 mm2, TPE, 0.5 m

Carrier, PA 6.6

Sanitary 2.5" clamp EN 1.4301 (AISI 304) 2
Type N Clamp EN 1.4301 (AISI 304) 2
Cable 4×21 AWG PUR, gray 10 m multistrand, with ferrules

Flame-retardant acc. to IEC 60332-1-2, FT1, VW1

Weight 2.7 kg (5.95 lb)

Electrical specifications

PR53 series inputs and outputs
Property Specification
Supply
Operating voltage 24 V DC nominal (9–30 V DC)
Power consumption Less than 1 W
Protection class 3, PELV
Outputs
Output parameters RI, temperature, concentration, quality factor
Diagnostics

Image detection: outside light error, no optical image, no sample, prism coated, prism coated persistently, outside light to prism, low image quality

Process temperature measurement: temperature measurement fault

Concentration calculation: concentration out of range, temperature out of range

Internal environment: high internal temperature, high internal humidity

System: blank image corrupted

Analog outputs
mA

Sourcing, isolated, NAMUR NE 43, configurable

mA range 3.8–20.5 mA
Loop impedance Max. 600 Ω
Accuracy of analog outputs at +20 °C ±0.1 % of full scale (±0.00002 RI)
Digital outputs
Digital output RS-485, non-isolated
Maximum cable run 300 m (approx. 1000 ft) (digital)
Supported protocol Modbus RTU
Connectors
External connectors

1 × M12 M 4 pins, A-coded 5

2 × M16×1.5 cable gland, Cable D 5–10 mm / Adapter for conduit entry M16×1.5 6/ NPT ½"

Environmental specifications

PR53AC operating environment
Property Specification
Process parameters
Process temperature –40 … +150 °C (–40 … +302 °F) 7
Design temperature +180 °C (356 °F) 8
Design pressure 40 bar 9
Operating environment

Storage temperature

−40 … +65 °C (−40 … +149 °F)

Operating temperature

−40 … +60 °C (−40 … +140 °F)

Maximum operating altitude

2000 m (approx. 6500 ft)

Operating humidity

0–100 %RH

Storage humidity

0–100 %RH, non‑condensing

UL 50E (NEMA) rating

Type 4X: Dust-tight. Protected from corrosion and hose-directed water.

IP rating

IP66: Dust-tight. Protected from powerful water jets from any direction.

IP67: Dust-tight. Protected from the effects of temporary immersion in water under standardized conditions of pressure and time.

1 EN 10204 / 3.1 certificate included.
2 Manufacturer's declaration included.
3 ADI free, FDA 21 C.F.R 177.1550, 3A Sanitary Standard, USP Class VI <88>, 70 °C.
4 3-A certificate, EHEDG certificate.
5 For USB2 adapter and Insight software, see www.vaisala.com/insight.
6 Conduit hub is not compatible with PR53 Safe-Drive system
7 –40 ... +130 °C (-40 ... 266 °F) EPDM gasket, -40... +150 °C (–40 … +302 °F) PTFE gasket.
8 Maximum momentary temperature peak.
9 Maximum at +20 °C (68 °F), operating pressure to the clamp rating pressure.