Mechanical specifications
Property | Specification |
---|---|
Wetted parts | |
Sensor head | EN 1.4435 BN2 (AISI 316L) 1 |
Surface roughness | Ra 0.8 μm |
Prism | Sapphire monocrystalline, 99.996 % Al2O3 2 |
Prism gasket | Modified PTFE 3 |
Sanitary 2.5" gasket |
EPDM 2 |
Type N gasket | EPDM 2 |
Welding ferrule | EN 1.4435 (AISI 316L) 1
4 ASME BPE-2019 (DIN 32676-C) |
Non-wetted parts | |
Housing | EN 1.4404 (AISI 316L) |
Screws TX20, torque 2.0 Nm | EN 1.4404 (AISI 316L) |
Cable gland |
EN 1.4305 (AISI 303) HUMMEL 1.693.1600.50 |
Dummy plug |
EN 1.4305 (AISI 303) AGRO 8717.96.08.70 |
Conduit hub |
EN 1.4404 (AISI 316L) Vaisala, DRW257718, M16×1.5 / NPT ½ in |
M12 connector |
Gland, EN 1.4305 (AISI 303) Contacts, CuZn with Ni/Au plating Phoenix Contact, 1405233, M12/4(M), A, 4×0.34 mm2, TPE, 0.5 m Carrier, PA 6.6 |
Sanitary 2.5" clamp | EN 1.4301 (AISI 304) 2 |
Type N Clamp | EN 1.4301 (AISI 304) 2 |
Cable | 4×21 AWG PUR, gray 10 m multistrand, with ferrules Flame-retardant acc. to IEC 60332-1-2, FT1, VW1 |
Weight | 2.7 kg (5.95 lb) |
Electrical specifications
Property | Specification |
---|---|
Supply | |
Operating voltage | 24 V DC nominal (9–30 V DC) |
Power consumption | Less than 1 W |
Protection class | 3, PELV |
Outputs | |
Output parameters | RI, temperature, concentration, quality factor |
Diagnostics |
Image detection: outside light error, no optical image, no sample, prism coated, prism coated persistently, outside light to prism, low image quality Process temperature measurement: temperature measurement fault Concentration calculation: concentration out of range, temperature out of range Internal environment: high internal temperature, high internal humidity System: blank image corrupted |
Analog outputs | |
mA |
Sourcing, isolated, NAMUR NE 43, configurable |
mA range | 3.8–20.5 mA |
Loop impedance | Max. 600 Ω |
Accuracy of analog outputs at +20 °C | ±0.1 % of full scale (±0.00002 RI) |
Digital outputs | |
Digital output | RS-485, non-isolated |
Maximum cable run | 300 m (approx. 1000 ft) (digital) |
Supported protocol | Modbus RTU |
Connectors | |
External connectors |
1 × M12 M 4 pins, A-coded 5 2 × M16×1.5 cable gland, Cable D 5–10 mm / Adapter for conduit entry M16×1.5 6/ NPT ½" |
Environmental specifications
Property | Specification |
---|---|
Process parameters | |
Process temperature | –40 … +150 °C (–40 … +302 °F) 7 |
Design temperature | +180 °C (356 °F) 8 |
Design pressure | 40 bar 9 |
Operating environment | |
Storage temperature |
−40 … +65 °C (−40 … +149 °F) |
Operating temperature |
−40 … +60 °C (−40 … +140 °F) |
Maximum operating altitude |
2000 m (approx. 6500 ft) |
Operating humidity |
0–100 %RH |
Storage humidity |
0–100 %RH, non‑condensing |
UL 50E (NEMA) rating |
Type 4X: Dust-tight. Protected from corrosion and hose-directed water. |
IP rating |
IP66: Dust-tight. Protected from powerful water jets from any direction. IP67: Dust-tight. Protected from the effects of temporary immersion in water under standardized conditions of pressure and time. |