Measurement type | Probe models |
---|---|
Humidity and temperature | HMP1, HMP3, HMP4, HMP5, HMP7, HMP8, HMP9 |
Temperature | TMP1 |
Dew point | DMP5, DMP6, DMP7, DMP8 |
Carbon dioxide | GMP251, GMP252 |
Vaporized hydrogen peroxide | HPP271, HPP272 |
Moisture in oil | MMP8 |
Property | Description/Value |
---|---|
Operating temperature |
With display: −20 … +60 °C (−4 …+140 °F) Without display: −40 … +60 °C (−40 …+140 °F) |
Storage temperature |
With display: −30 … +70 °C (−22 …+158 °F) Without display: −40 … +70 °C (−40 …+158 °F) |
Operating humidity |
0–100 %RH |
Maximum operating altitude |
5000 m (approx. 16 400 ft) |
IP rating |
IP66 1
|
Property | Description/Value |
---|---|
Power supply input |
15–30 V DC 2 24 V AC ±10 % 50/60 Hz |
Fuse size for power supply | 2.5 A |
Transmitter service port connection |
|
Analog outputs | Current or voltage |
Number of analog outputs | 3 |
Isolation | Not galvanically isolated |
Selectable voltage output types | 0–1 V, 0–5 V, 0–10 V, 1– 5 V, scalable, RL ≥ 10 kΩ |
Selectable current output types | 4–20 mA, 0–20 mA, scalable, RL ≤ 500 Ω |
Accuracy of analog outputs at 20 °C (+68 °F) | ±0.1 % full scale 4 |
Temperature dependence | ±0.005 % / °C full scale |
Current consumption at 20 °C (+68 °F) (Uin 24 V DC) | |
Minimum consumption with display off, no analog outputs active, no probe connected 5 | 13 mA |
Minimum consumption with display on, brightness normal mode, no analog outputs active, no probe connected | 18 mA |
Uout 0–1 V, 0–5 V, 0–10 V, 1–5 V | + 1.8 mA per connected channel at maximum load |
Iout 4–20 mA, 0–20 mA | + 12.3 mA max. per connected channel |
Property | Description/Value |
---|---|
EU directives and regulations |
EMC Directive (2014/30/EU) RoHS Directive (2011/65/EU) as amended by 2015/863 |
Electromagnetic compatibility (EMC) |
EN IEC 61326-1, industrial environment |
EMC emissions |
CISPR 32 / EN 55032, Class A FCC part 15 B, Class A ICES‑3 / NMB‑3 (Class A) |
Compliance marks |
CE, China RoHS, FCC, KC, RCM, UKCA |
Property | Description/Value |
---|---|
Housing material | EN AW-6082 |
Connection screw terminals | Max. 1.5 mm2 wire (16 AWG) |
Cable lead-throughs for output and power cables |
|
Cable diameter for M20×1.5 gland | 7–13 mm (0.26–0.51 in) |
Cable diameter for M16×1.5 gland | 2–6 mm (0.08–0.24 in) |
Dimensions |
161 × 134 × 43.5 mm (6.34 × 5.26 × 1.71 in) |
Weight | 1200 g (2.65 lb) |
IP40 for transmitters equipped with a locking wheel and a probe connection cable attached to it. This combination is recommended only for indoor conditions.