Specifications - INDIGO300

Indigo300 User Guide

Document code
M212849EN
Revision
C
ft:locale
en-US
Product
INDIGO300
Document type
User guide
Indigo-compatible probes
Measurement type Probe models
Humidity and temperature HMP1, HMP3, HMP4, HMP5, HMP7, HMP8, HMP9
Temperature TMP1
Dew point DMP5, DMP6, DMP7, DMP8
Carbon dioxide GMP251, GMP252
Vaporized hydrogen peroxide HPP271, HPP272
Moisture in oil MMP8
Indigo300 operating environment
Property Description/Value

Operating temperature

With display: −20 … +60 °C (−4 …+140 °F)

Without display: −40 … +60 °C (−40 …+140 °F)

Storage temperature

With display: −30 … +70 °C (−22 …+158 °F)

Without display: −40 … +70 °C (−40 …+158 °F)

Operating humidity

0–100 %RH

Maximum operating altitude

5000 m (approx. 16 400 ft)

IP rating

IP66 1
Indigo300 inputs and outputs
Property Description/Value
Power supply input

15–30 V DC 2

24 V AC ±10 % 50/60 Hz

Fuse size for power supply 2.5 A
Transmitter service port connection
  • Connection to Insight with USB2 and cable 262195SP 3
  • Connection to Indigo80 with cable 262195SP
Analog outputs Current or voltage
Number of analog outputs 3
Isolation Not galvanically isolated
Selectable voltage output types 0–1 V, 0–5 V, 0–10 V, 1– 5 V, scalable, RL ≥ 10 kΩ
Selectable current output types 4–20 mA, 0–20 mA, scalable, RL ≤ 500 Ω
Accuracy of analog outputs at 20 °C (+68 °F) ±0.1 % full scale 4
Temperature dependence ±0.005 % / °C full scale
Current consumption at 20 °C (+68 °F) (Uin 24 V DC)
Minimum consumption with display off, no analog outputs active, no probe connected 5 13 mA
Minimum consumption with display on, brightness normal mode, no analog outputs active, no probe connected 18 mA
Uout 0–1 V, 0–5 V, 0–10 V, 1–5 V + 1.8 mA per connected channel at maximum load
Iout 4–20 mA, 0–20 mA + 12.3 mA max. per connected channel
Indigo300 compliance
Property Description/Value

EU directives and regulations

EMC Directive (2014/30/EU)

RoHS Directive (2011/65/EU) as amended by 2015/863

Electromagnetic compatibility (EMC)

EN IEC 61326-1, industrial environment

EMC emissions

CISPR 32 / EN 55032, Class A

FCC part 15 B, Class A

ICES‑3 / NMB‑3 (Class A)

Compliance marks

CE, China RoHS, FCC, KC, RCM, UKCA
Indigo300 mechanical specifications
Property Description/Value
Housing material EN AW-6082
Connection screw terminals Max. 1.5 mm2 wire (16 AWG)
Cable lead-throughs for output and power cables
  • M20×1.5 cable gland / conduit fitting NPT 1/2"
  • M16×1.5 cable gland / conduit fitting NPT 1/2"
Cable diameter for M20×1.5 gland 7–13 mm (0.26–0.51 in)
Cable diameter for M16×1.5 gland 2–6 mm (0.08–0.24 in)
Dimensions

161 × 134 × 43.5 mm (6.34 × 5.26 × 1.71 in)

Weight 1200 g (2.65 lb)
1

IP40 for transmitters equipped with a locking wheel and a probe connection cable attached to it. This combination is recommended only for indoor conditions.

2 When used with the HMP7 probe, the minimum required power supply input is 18 V DC.
3 Vaisala Insight software for Windows® available at vaisala.com/insight.
4 For the voltage outputs, small variation is possible around true zero.
5 For the current consumption of the connected probe, see the probe's user documentation at docs.vaisala.com.