This example shows how to use Indigo80 to
configure the transmitter's analog output 1 to use the following settings when a humidity
and temperature probe is connected to the transmitter:
- 4 ... 20 mA current output
- Relative humidity measurement scaled to
0–100 %RH
- Clipping limit 5 %, error limit 10 %, and
error output 3.6 mA
-
Connect the transmitter to the indicator.
-
Open the indicator main menu by pressing
.
-
Select Devices. If you have more than one device connected to the indicator, make a further selection
between the devices.
In this case, select Indigo300.
-
Select Settings.
-
In Analog output type, select Current output.
-
Indigo80 reloads the transmitter. Reselect .
-
Scroll down and select Analog output 1.
-
In Output mode, select 4...20 mA.
-
In Output parameter, select Relative humidity.
-
Select Scale low end. Use the arrow buttons to set the value as 0.00, and then select
Set.
-
Select Scale high end. Use the arrow buttons to set the value as 100.00, and then
select Set.
-
Select Error output level. Use the arrow buttons to set the value as 3.60, and then select
Set.
-
Select Clipping limit. Use the arrow buttons to set the value as 5.0, and then select
Set.
-
Select Error limit. Use the arrow buttons to set the value as 10.0, and then select
Set.
Your configuration is now in use.
-
Exit the menu
by pressing
or return to the main menu by pressing
.