Clearing transmitter analog output configuration - INDIGO300

Indigo300 User Guide

Document code
M212849EN
Revision
B
Language
English
Product
INDIGO300
Document type
User guide

If you need to restore the transmitter to a state where no analog output configuration exists in the transmitter, you can clear the current configuration with the Indigo80 handheld indicator.

When you remove the analog output configuration from the transmitter, the transmitter automatically adapts the analog output settings of the next probe you connect to it.
  1. Connect the transmitter to the indicator.
  2. Open the indicator main menu by pressing .
  3. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.

    In this case, select Indigo300.

  4. Select Reset analog output settings > OK.
  5. Confirm by selecting OK.
  6. Exit the menu by pressing or return to the main menu by pressing .