Wiring options - INDIGO300

Indigo300 Quick Guide

Document code
M212848EN
Revision
B
Language
English
Product
INDIGO300
Document type
User guide
CAUTION Do not energize the power supply before the wiring has been connected.

The transmitter wiring options include a lead-through at the backside and lead-throughs on the bottom.

Figure 1. Indigo300 wiring options and cable diameters
1
Wiring from the back
  • Cable diameter: 5 … 10 mm (0.20 … 0.39 in)
2
Wiring through a M20×1.5 gland or conduit fitting NPT 1/2"
  • Cable diameter for M20 gland: 7 … 13 mm (0.26 … 0.51 in)
3
Wiring through a M16×1.5 gland or conduit fitting NPT 1/2"
  • Cable diameter for M16 gland: 2 … 6 mm (0.08 … 0.24 in)
If you are using a 24 V AC power supply for the transmitter, make sure to use separate cables for the AC power and the analog outputs to ensure that the EMC requirements are fulfilled.