USB connection to Insight PC software cannot be established |
USB connection has been disabled with the DIP switch on the transmitter component
board. |
Enable the USB connection with the DIP switch. |
Functionalities are not available in Insight PC software. |
Insight PC software is in Basic
mode. Certain functionalities are only available in Advanced mode. |
Change the user mode from Basic
mode to Advanced mode in the Settings menu of Insight PC software. |
Measurement performance affected after calibration. |
Probe has not been switched from calibration mode to normal operating mode in the
Insight PC software. |
Open the Calibrate menu in Insight PC software. Exit the menu with the Close selection, and ensure that you confirm closing the calibration mode when exiting
with the Yes selection. |